PayoutCanada
← All listings
Product RecallProduct RecallHighRecall Date: Jul 23, 2026

OverStitch™ Suture Cinch and Suture Cinch–Long

Boston Scientific is initiating removal of specific lots of OverStitch™ Suture Cinch and Suture Cinch-Long (collectively "Suture Cinch"). these devices are accessories used with the OverStitch™ Endoscopic Suturing System, OverStitch™ NXT Endoscopic Suturing System, and the X-Tack™ Endoscopic HeliX Tacking System to secure and cut the proximal end of a suture through the working channel of a gastroscope or colonoscope.During device activation, the Suture Cinch may fail to deploy the cinch implant and/or fail to cut the suture as intended. The issue cannot be detected through pre-use inspection but becomes apparent once the user attempts to deploy the device during the procedure.

Status
Active
Key date
Jul 23, 2026
Provinces
National
Added
Jul 24, 2026

Take action

Confirm your eligibility and the exact deadline on the official page, then file directly with the administrator. PayoutCanada does not file on your behalf.