Product RecallProduct RecallHighRecall Date: Jul 16, 2026
V-Loc™ 90 and 180 Suture Products
On 29-Jan-2026 during a quality inspection at a manufacturing site, a V-Loc 180 unit was found with a hole in the breather pouch on the Tyvek side. The breather pouch serves as the sterile barrier system for the suture. Investigation determined the briller plate on the sealing station may be causing the hole. No complaints have been identified for this issue, but lots within the issue bounding have been released to Distribution Centers (DCs) and sold to customers. These released lots may also contain units with holes in the Tyvek side of the breather pouch.
- Status
- Active
- Key date
- Jul 16, 2026
- Provinces
- National
- Added
- Jul 17, 2026
Take action
Confirm your eligibility and the exact deadline on the official page, then file directly with the administrator. PayoutCanada does not file on your behalf.
View official source & file →