PayoutCanada
← All listings
Product RecallProduct RecallHighRecall Date: Aug 20, 2026

Patient Information Centre iX

Philips received reports that some PIC iX systems were frozen and could not be used. It was determined that failure of the watchdog may cause the system to be in an unknown state and therefore potentially prevent alarms from being generated or logged. Following the initial system error which triggers the watchdog, staff may observe unexpected system behavior while the watchdog attempts recovery. The observed behavior depends on the underlying software issue and may include:-Inability to interact with the PIC iX user interface. If real-time waveforms, numeric data,and alarms remain operational, there is no impact to bedside or central monitoring and alarming.-The user will not recognize the device issue before the hazardous situation occurs if theobserved behavior is: Loss of central alarming, Loss of, or frozen, waveforms/data display-Bedside monitors and telemetry devices could generate a "No Central Mon" INOP

Status
Active
Key date
Aug 20, 2026
Provinces
National
Added
Aug 21, 2026

Take action

Confirm your eligibility and the exact deadline on the official page, then file directly with the administrator. PayoutCanada does not file on your behalf.