Product RecallProduct RecallHighRecall Date: Aug 28, 2026
Blanketrol® II 115V and Blanketrol® III 115V
Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available.
- Status
- Active
- Key date
- Aug 28, 2026
- Provinces
- National
- Added
- Aug 31, 2026
Take action
Confirm your eligibility and the exact deadline on the official page, then file directly with the administrator. PayoutCanada does not file on your behalf.