PayoutCanada
← All listings
Product RecallProduct RecallHighRecall Date: Oct 15, 2026

cobas® c 303 analytical unit

An issue has been identified where vacuum tubing of the cell rinse unit of the cobas c 303 analytical unit prematurely deteriorated. Investigations confirmed the issue. In the event of small pinholes or cracks, no impact on analytical performance was identified. However, in case of tubing rupture, the following can occur as a result of the issue: Undetected leakage: Fluid may drip into random reaction cells. Result impact: Increased risk of erroneous patient and QC results.

Status
Active
Key date
Oct 15, 2026
Provinces
National
Added
Oct 15, 2026

Take action

Confirm your eligibility and the exact deadline on the official page, then file directly with the administrator. PayoutCanada does not file on your behalf.

View official source & file →